International Journal of Medical and Pharmaceutical Research
2026, Volume-7, Issue 2 : 251-255
Research Article
Comparison Of the Clinical Performance of Two Second-Generation Supraglottic Airway Devices (I-Gel vs Ambu Auragain) In Adult Patients Undergoing Elective Surgery Under General Anesthesia
 ,
 ,
 ,
 ,
Received
Feb. 3, 2026
Accepted
Feb. 24, 2026
Published
March 8, 2026
Abstract

Background: Second-generation supraglottic airway devices are increasingly used as alternatives to endotracheal intubation because of their ease of insertion and improved airway sealing characteristics.

Aim: To compare the clinical performance of I-gel and Ambu AuraGain in adult patients undergoing elective surgery under general anaesthesia.

Methods: A prospective randomized observational study was conducted in 60 patients divided into two groups receiving either I-gel or Ambu AuraGain. Quality of insertion, oropharyngeal leak pressure, airway sealing quality score, gastric tube insertion, hemodynamic parameters, and postoperative complications were evaluated.

Results: Demographic characteristics were comparable between groups. Ease of insertion and number of attempts were similar; however, insertion time was significantly shorter with I-gel (15.87±1.36 s) compared to AuraGain (20±2.81 s). Oropharyngeal leak pressure was significantly higher in the AuraGain group (28.33±1.52 cmH₂O vs 22.57±1.01 cmH₂O). Gastric tube insertion required fewer attempts with AuraGain, while hemodynamic parameters and postoperative complications were comparable.

Conclusion: Both devices were effective and safe for airway management; I-gel provided faster insertion whereas Ambu AuraGain offered superior airway sealing and gastric tube insertion performance

Keywords
INTRODUCTION

Secure airway management is a cornerstone of safe anaesthesia practice, particularly during elective surgical procedures requiring general anaesthesia. Traditionally, endotracheal intubation has been the gold standard for airway control; however, supraglottic airway devices (SADs) offer a less invasive alternative that can reduce airway trauma and sympathetic responses while maintaining adequate ventilation and oxygenation. Second-generation SADs, such as the i-gel® and Ambu® AuraGain™, are designed with features including improved sealing pressures, integrated gastric channels, and enhanced conduits for intubation, which aim to increase safety and clinical performance compared to earlier SAD models [1,2].

 

The i-gel® is a single-use, non-inflatable cuffed SAD made of a thermoplastic elastomer that conforms to the perilaryngeal anatomy for a secure seal, eliminating the need for cuff inflation and potentially reducing placement time [3,4]. In contrast, Ambu® AuraGain™ features an anatomically curved design, a gastric access port, and the ability to act as a conduit for fibreoptic-guided intubation, positioning it as a versatile second-generation device in airway management [5,6]. Both devices have demonstrated acceptable first-attempt success rates and adequate oropharyngeal leak pressures (OLP), though differences in insertion characteristics and seal efficiency remain subjects of ongoing clinical investigation.

 

Several randomized trials and clinical studies comparing these devices have highlighted subtle performance variations. For example, adult studies show that both i-gel and AuraGain safely facilitate ventilation and preserve hemodynamic stability during elective procedures, with comparable ease of placement and perioperative complications [7]. Some trials indicate that i-gel may have a shorter insertion time and lower postoperative airway morbidity, while AuraGain may offer higher oropharyngeal seal pressures favorable for positive-pressure ventilation [8,9]. Additionally, evidence suggests that these devices perform reliably across diverse surgical populations, including obese patients and during laparoscopic procedures, further supporting their expanding role in modern anaesthesia practice [10].

 

Despite these insights, variability in reported outcomes underscores the need for continued comparative research to clarify the clinical performance differences between i-gel and Ambu AuraGain, forming the rationale for the current study.

 

MATERIAL AND METHODS

This prospective randomized observational study was conducted after obtaining approval from the Institutional Review Board to compare the clinical performance of two second-generation supraglottic airway devices, I-gel and Ambu AuraGain, in adult patients undergoing elective surgical procedures under general anaesthesia. A total of 60 patients of either sex, aged between 18 and 60 years and belonging to American Society of Anaesthesiologists (ASA) physical status grade I and II, were included in the study. Patients scheduled for elective surgeries with adequate mouth opening and suitable airway anatomy were enrolled after obtaining informed written consent. Patients with mouth opening less than 2.5 cm, full stomach, gastro-oesophageal reflux disease, hiatus hernia, cervical spine disease, or pharyngeal and laryngeal pathology were excluded. Pre-anaesthetic evaluation included detailed medical and surgical history, airway assessment using Mallampati grading, general and systemic examination, and routine investigations such as complete blood count, renal and liver function tests, serum electrolytes, chest X-ray, and electrocardiogram.

 

On the day of surgery, standard monitoring including electrocardiography, non-invasive blood pressure, pulse oximetry, and end-tidal carbon dioxide was applied, and baseline vital parameters were recorded. Patients were randomly allocated into two groups of 30 each using an odd–even allocation method, with Group I receiving the I-gel and Group A receiving the Ambu AuraGain. Device size selection was based on the patient’s body weight according to manufacturer recommendations. All patients were kept nil per oral for at least six hours prior to surgery. Premedication consisted of intravenous glycopyrrolate 0.004 mg/kg, ondansetron 4 mg, and fentanyl 2 µg/kg. After pre-oxygenation with 100% oxygen for three minutes, induction of anaesthesia was performed using intravenous propofol 2.5 mg/kg, with additional doses administered if required to achieve adequate depth of anaesthesia.

 

Following loss of eyelash reflex and adequate jaw relaxation, the assigned supraglottic airway device was inserted using standard insertion techniques. In Group A, the cuff of the Ambu AuraGain was inflated according to size recommendations, whereas in the Group I, I-gel was inserted without cuff inflation. Proper placement was confirmed by bilateral chest expansion, square-wave capnography, and absence of audible leak. Anaesthesia was maintained using a mixture of oxygen, nitrous oxide, and sevoflurane with intermittent positive pressure ventilation, along with intravenous atracurium for muscle relaxation. Quality of insertion was assessed based on ease of insertion, number of attempts, and time required for successful airway placement. Working performance was evaluated by measuring oropharyngeal leak pressure and airway sealing quality score based on tidal volume loss.

 

Airway sealing quality was graded by ASQS as determined by percentage loss of delivered tidal volume (inspiratory (set) – expiratory (outcome) volume on ventilator display screen) [11].

 

Airway sealing quality score

1

No leak detected

2

Minor leak of tidal volume (Vt loss <20%)

3

Moderate leak of tidal volume (Vt loss 20% - 40%)

4

Insufficient seal (Vt loss >40%)

 

Complications during insertion such as coughing, laryngospasm, bronchospasm, and regurgitation were recorded. Gastric tube insertion was attempted after airway placement, and its ease and number of attempts were documented. Hemodynamic parameters including pulse rate, systolic and diastolic blood pressure, mean arterial pressure, oxygen saturation, and end-tidal carbon dioxide were recorded at predefined intervals before induction and after insertion of the airway device. At the end of surgery, neuromuscular blockade was reversed using neostigmine and glycopyrrolate, and the airway device was removed after adequate recovery. Post-operative complications including sore throat, blood-stained device, hoarseness of voice, and airway trauma were noted. Statistical analysis was performed using SPSS software, with continuous variables expressed as mean ± standard deviation. The unpaired t-test and chi-square test were used for comparison, and a p-value less than 0.05 was considered statistically significant.

 

RESULTS

Table 1 show the demographic characteristics of the study population, where both groups were comparable in baseline parameters. The mean age was 34 years in Group I and 32.1 years in Group A. Gender distribution was similar between both the groups. Mean body weight was identical at 60.5 kg in both groups, while ASA grade I/II distribution was 22/8 and 23/7 respectively, confirming comparable patient characteristics. Mallampati grading was also similar between the groups (20/10 vs 21/9), indicating homogeneity of airway assessment.

 

Table 2 illustrates the quality of insertion of both supraglottic airway devices. Easy insertion was achieved in 27 patients in Group I and 26 patients in Group A, showing no significant difference (p=0.687). First attempt success rate was high in both groups, with 28 patients in Group I and 29 patients in Group A (p=0.553). However, the time taken for insertion was significantly shorter in Group I (15.87±1.36 seconds) compared to Group A (20±2.81 seconds) with a highly significant p value, demonstrating faster placement with I-gel.

 

Table 3 presents the working performance outcomes. The mean oropharyngeal leak pressure was higher in Group A (28.33±1.52 cmH2O) than Group I (22.57±1.01 cmH2O), with a highly significant p value. Airway sealing quality score showed score 1 in 24 patients of Group I and 27 patients of Group A, while score 2 was seen in 6 patients and 3 patients respectively (p=0.278), indicating comparable sealing efficiency.

 

Table 4 shows gastric tube insertion characteristics. Easy insertion was observed in 20 patients in Group I and 25 patients in Group A (p=0.136). First attempt success for gastric tube insertion was higher in Group A (25 patients, 83.3%) compared to Group I (17 patients, 56.7%), which was statistically significant (p=0.024), indicating smoother gastric tube placement with Ambu AuraGain.

 

Table 5 demonstrates post-operative complications. Blood-stained device was observed in 1 patient (3.33%) in Group I and 3 patients (10%) in Group A, while sore throat occurred in 2 patients (6.67%) and 4 patients (13.3%) respectively. There were no cases of tongue-lip-dental trauma, hoarseness of voice, or bronchospasm/laryngospasm in either group, and the overall comparison was statistically non-significant (p=0.778), suggesting comparable safety profiles.

 

Table 1 - Demographic data

Patient data

Group I n=30

Group A n=30

Number of patients

30

30

Age (years)

34±12.97

32.1±8.48

Gender (Male/Female)

7/23

6/24

Weight (kg)

60.5±13.78

60.5±10.5

ASA grade(I/II)

22/8

23/7

MPG(I/II)

20/10

21/9

 

Table 2 - Quality of insertion

Quality of insertion

Group I n=30(%)

Group A n=30(%)

P value

Ease of insertion Easy

27(90%)

26(86.7%)

0.687

Difficult

3(10%)

4(13.3%)

 

No. of attempts First

28(93.3%)

29(96.7%)

0.553

Second

2(6.7%)

1(3.3%)

 

Failed

0

0

 

Time taken for insertion (seconds)

15.87±1.36

20±2.81

<0.00001

 

Table 3- Working performance

Working performance

Group I n=30 (%)

Group A n=30 (%)

P value

Oropharyngeal leak pressure (OLP)(in cmH2O)

22.57±1.01

28.33±1.52

0.00001

Airway sealing quality score 1

24(80%)

27(90%)

0.278

2

6(20%)

3(10%)

 

3

0

0

 

4

0

0

 

 

Table 4 - Gastric tube insertion

Gastric tube insertion

Group I n=30(%)

Group A n=30(%)

P value

Ease Easy

20(66.7%)

25(83.3%)

0.136

Difficult

10(33.3%)

5(16.7%)

 

Attempts 1

17(56.7%)

25(83.3%)

0.024

2

13(43.3%)

5(16.6%)

 

Failed

0

0

 

 

Table 5 - Post-operative complications

Post-operative Complications

Group I (n=30)

Group A (n=30)

P value

Blood-stained device

1 (3.33%)

3 (10%)

0.778

Tongue-lip-dental trauma

0

0

 

Sore throat

2 (6.67%)

4 (13.3%)

 

Hoarseness of voice

0

0

 

Bronchospasm/Laryngospasm

0

0

 

DISCUSSION
The present study compared two second-generation supraglottic airway devices, I-gel and Ambu AuraGain, with respect to quality of insertion, working performance, gastric tube insertion, hemodynamic stability and postoperative complications. The demographic variables between both groups were comparable, indicating proper randomization and eliminating confounding baseline differences. Similar findings were reported by Brain AIJ et al. [12], who described comparable patient characteristics in studies evaluating supraglottic airway devices, supporting the validity of randomized allocation methods in airway trials. In the current study, ease of insertion was higher in the I-gel group (90%) compared to the AuraGain group (86.7%), although the difference was statistically non-significant. These findings correlate with the observations of Kannaujia A et al. [13], who noted that the non-inflatable cuff and anatomical design of I-gel facilitate smoother insertion with minimal manipulation. Furthermore, shorter insertion time in the I-gel group (15.87±1.36 seconds vs 20±2.81 seconds) is consistent with the results reported by Pratheeba N et al. [14], where cuffless SADs demonstrated reduced insertion duration due to elimination of cuff inflation steps.

 

The working performance of both devices showed significant differences, particularly in oropharyngeal leak pressure (OLP), which was markedly higher in the AuraGain group (28.33±1.52 cmH₂O) compared to the I-gel group (22.57±1.01 cmH₂O). Higher OLP reflects superior airway sealing capability and suitability for positive pressure ventilation, a finding also emphasized by Wong DT et al. [15], who reported improved sealing pressures with anatomically curved second-generation SADs. Although airway sealing quality score (ASQS) was slightly better in the AuraGain group, the difference remained statistically non-significant, suggesting both devices provide effective ventilation in elective surgical settings. The improved seal with AuraGain may be attributed to its inflatable cuff and wider airway tube, which enhances contact with peri-laryngeal structures and reduces gas leak.

 

Gastric tube insertion was easier and required fewer attempts in the AuraGain group, with first-attempt success of 83.3% compared to 56.7% in the I-gel group. This finding aligns with observations by Mihara T et al. [16], who demonstrated that devices with larger gastric channels facilitate smoother passage of gastric tubes and reduce manipulation. The presence of a dedicated low-friction gastric access port in AuraGain likely contributed to improved success rates. Hemodynamic parameters including pulse rate, systolic blood pressure, diastolic blood pressure, and mean arterial pressure remained comparable between groups throughout the perioperative period, indicating minimal sympathetic stimulation during insertion of either device. These findings reinforce previous literature suggesting that supraglottic airway devices provide stable hemodynamic profiles compared to traditional airway techniques.

 

Post-operative complications such as sore throat and blood-stained device were slightly more frequent in the AuraGain group but did not reach statistical significance. This may be explained by the inflatable cuff exerting higher mucosal pressure compared to the gel-based non-inflatable design of I-gel. Overall, both devices demonstrated high success rates, effective ventilation, and acceptable safety profiles, supporting their clinical utility in adult elective surgeries. The present findings highlight that while I-gel offers faster insertion and ease of placement, Ambu AuraGain provides superior airway seal and improved gastric tube insertion performance, thereby offering distinct advantages depending on surgical requirements and anesthesiologist preference.

 

CONCLUSION:
Both second-generation supraglottic airway devices, I-gel and Ambu AuraGain, were effective in maintaining airway patency and ventilation in adult patients undergoing elective surgery under general anaesthesia. I-gel demonstrated shorter insertion time and slightly easier placement, whereas Ambu AuraGain provided significantly higher oropharyngeal leak pressure and improved gastric tube insertion characteristics. Hemodynamic stability and postoperative complications were comparable between the groups. Therefore, Ambu AuraGain may be considered a preferable alternative when higher airway sealing pressure is required, while I-gel remains advantageous for rapid and atraumatic airway placement.

 

REFERENCES

  1. Anand Kuppusamy A, Guruswamy J, Parasakthi K. Comparison of Proseal laryngeal mask airway (Proseal LMA) with I-gel by fibreoptic view of glottis aperture in anaesthetised adult patients. Indian J Anaesth. 2016;3(1):12-15.
  2. European Resuscitation Council, American Heart Association, International Liaison Committee on Resuscitation. International guidelines for cardiopulmonary resuscitation and emergency cardiac care: an international consensus on science. Resuscitation. 2000;6:29-71.
  3. Peppard SB, Dickens JH. Laryngeal injury following short-term intubation. Ann Otol Rhinol Laryngol. 1983;92:327-330.
  4. Cork RC, Depa RM, Standen JR. Prospective comparison of use of the laryngeal mask and endotracheal tube for ambulatory surgery. Anesth Analg. 1994;79:719-727.
  5. Miller RD. Miller’s Anesthesia. 8th ed. Philadelphia: Elsevier; 2015.
  6. Jayashree S. Laryngeal mask airway and its variants. Indian J Anaesth. 2005;49:275-280.
  7. Griffin RM, Hatcher IS. Aspiration pneumonia and the laryngeal mask airway. Anaesthesia. 1990;45:1039-1040.
  8. Malshetwar KV, Totawar SR. Comparative study of supraglottic airway device Proseal versus I-gel in mechanically ventilated patients. MedPulse Int J Anaesthesiol. 2017;3(2):59-65.
  9. Singh K, Gurha P. Comparative evaluation of Ambu AuraGain with ProSeal laryngeal mask airway in patients undergoing laparoscopic cholecystectomy. Indian J Anaesth. 2017;61(6):469-474.
  10. Nasir MA. I-gel user guide. 7th ed. Wokingham: Intersurgical Ltd; 2009.
  11. Youssef MMI, Dobal NM, Hammad YM, El-Refai NAR, Abdelrahman RAA. Comparison between Air-Q Self Pressurized Airway Device with Blocker and Proseal Laryngeal Mask Airway in anesthetized paralyzed adult female patients undergoing elective gynecological operations. Anaesthesiol Intensive Ther. 2024 Jul 23;56(2):108–20. doi: 10.5114/ait.2024.141203. PMCID: PMC11284585.
  12. Brain AIJ. The laryngeal mask: a new concept in airway management. Br J Anaesth. 1983;55:801-805.
  13. Kannaujia A, Srivastava U, Saraswat N, Mishra A, Kumar A, Saxena S. A preliminary study of I-gel: a new supraglottic airway device. Indian J Anaesth. 2009;53(1):52-56.
  14. Pratheeba N, Ramya GS, Ranjan RV, Remadevi R. Comparison of i-gel and laryngeal mask airway Classic in terms of ease of insertion and hemodynamic response: a randomized observational study. Anesth Essays Res. 2016;10(3):521-525.
  15. Wong DT, Ooi A, Singh KP, Dallaire A, Meliana V, Lau J, Chung F, Singh M, Wong J. Comparison of oropharyngeal leak pressure between Ambu AuraGain and LMA Supreme supraglottic airways: a randomized controlled trial. Can J Anaesth. 2018;65(7):797-805.
  16. Mihara T, Nakayama R, Ka K, Goto T. Comparison of the clinical performance of I-gel and Ambu AuraGain in children: a randomized noninferiority clinical trial. Eur J Anaesthesiol. 2019;36(6):411-417.
Recommended Articles
Research Article Open Access
Functional Outcome In Repair of Zone V Flexor Tendon Injuries in a Tertiary Care Centre in Kerala
2026, Volume-7, Issue 2 : 1714-1722
Research Article Open Access
Cadaveric Study of Branching Pattern and Clinically Relevant Variations of the Axillary Nerve
2026, Volume-7, Issue 2 : 1723-1736
Research Article Open Access
Incidence and risk factors of neonatal hypoglycaemia in term and preterm infants
2026, Volume-7, Issue 2 : 1747-1752
Research Article Open Access
Multiparametric MRI Evaluation of Cervical Cancer with Histopathological Correlation
2026, Volume-7, Issue 2 : 1706-1713
International Journal of Medical and Pharmaceutical Research journal thumbnail
Volume-7, Issue 2
Citations
76 Views
67 Downloads
Share this article
License
Copyright (c) International Journal of Medical and Pharmaceutical Research
Creative Commons Attribution License Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.
All papers should be submitted electronically. All submitted manuscripts must be original work that is not under submission at another journal or under consideration for publication in another form, such as a monograph or chapter of a book. Authors of submitted papers are obligated not to submit their paper for publication elsewhere until an editorial decision is rendered on their submission. Further, authors of accepted papers are prohibited from publishing the results in other publications that appear before the paper is published in the Journal unless they receive approval for doing so from the Editor-In-Chief.
IJMPR open access articles are licensed under a Creative Commons Attribution-ShareAlike 4.0 International License. This license lets the audience to give appropriate credit, provide a link to the license, and indicate if changes were made and if they remix, transform, or build upon the material, they must distribute contributions under the same license as the original.